Improved optical performance facilitates microscopic analysis of cells

Cell Culture Consumables

Cell Culture Consumables

Ines Kristina Hartmann, Wolf Wente, Linda Knauer, Khandaker Siddiquee, Ma Sha

Microscopic analysis of cells using an inverted microscope is a daily routine procedure in every cell culture laboratory. Visual examination of cultures is crucial e.g. to assess the degree of confluence of the cell layer and confirm the absence of contaminants. Furthermore a lot of cell culture based assays performed in plate formats are analyzed using manual or automated microscopic techniques. Besides instrumentation, the choice of the consumable strongly influences the microscopic results.

The daily analysis of cell cultures using an inverted microscope is an essential step in cell handling. From simple routine procedures – like the quick visual check of cell morphology or the assessment of confluence of the cell layer – to complex analysis of cell based assays, a quick and reliable optical examination is crucial for successful cell culture research. The following factors are considered as a prerequisite for efficient microscopic analysis: a minimum of refocusing across the area of observation, a uniform illumination without interfering shadows and a highcontrast image of the cells across the whole growth area. Besides the microscope itself, the consumable strongly influences the operational procedure and the imaging results in different ways:

Focus stability

Anzeige

The material, bottom thickness and geometry influence the planarity of a cell culture consumable. Glass for instance is a very rigid and inflexible material that is highly planar and very well suited for high resolution imaging. Polystyrene, the material cell culture consumables are made of, is more flexible resulting in a less planar surface in comparison to glass. A consumable with a poor bottom planarity may result in increased refocusing and can make manual microscopic analysis a time-consuming and cumbersome procedure. Furthermore the repetitive motion of the hand constantly moving the focus knob to readjust the microscope objective to the right focus level means a less ergonomic work situation. This can be tedious especially when analyzing a large amount of vessels on a daily basis or many well to well analyses in plate formats have to be performed. Using an automated microscope e.g. for high content analysis (HCA) of cells solves the ergonomic problem but the laser-based autofocus mechanism implies repeated refocusing of the lens resulting in prolonged analysis steps or even worse, in less reproducible results.

Meniscus formation and phase contrast performance

Phase contrast microscopy is the most common contrast enhancing light microscopic technique that is used for cell analysis as it creates high-contrast images of usually transparent specimens such as living cells. A common problem in phase contrast microscopy is the meniscus formation at the air-liquid interface. The curvature of the liquid causes a refraction of the light with the result that cells can hardly be displayed with a satisfactory contrast thus making it difficult to use this technique in smaller formats, like 96-well plates. There are specialized consumables available for meniscus correction [1] but using standard cell culture formats the meniscus is something that can hardly be avoided. In addition to its optical disadvantage – the loss of image quality – the meniscus is also associated with a phenomenon that is called the halo cell growth effect: cells often tend to adhere as a halo in dishes or plates in relation to the center. It has been suggested that the meniscus formation of the media leads to this irregular cell growth when seeding cells or even re-feeding cells with new media [2].

Illumination of the imaging area

Interfering shadows at the well edges are a common problem when working with multiwell plates limiting the area of observation. This can be a disadvantage especially in applications, where a uniform illumination of the whole growth area is crucial e.g. for single cell cloning experiments: here interfering shadows at the well edges make it more difficult or even impossible to identify cell clones growing at the periphery of the well.

In this Application Note we compare the optical performance of Eppendorf Cell Culture Consumables with corresponding products from other suppliers: We analyzed planarity, focus stability, phase contrast image quality and shadow formation.

Material and Methods

Planarity and focus stability

Whole plate planarity of different 96-well polystyrene cell culture plates was measured by optical contact scanning (n = 4). To analyze focus stability across the growth area NIH 3T3 cells were seeded at equal density in 100 mm Eppendorf Cell Culture Dishes, and comparable dishes from other suppliers (competitor A, B). Cells were analyzed 48 hours post seeding with an inverted microscope using a 10 x objective (Axio Observer A1, Zeiss®). Cells were focused in the center of the dish and then the dish was moved to different points of the growth area without refocusing. Pictures were taken at three different points of the dish (see drawing in figure 1) and it was compared if the cells can be seen in focus.

Phase contrast performance

To examine the meniscus effect in phase contrast inverted microscopy, CHO-K1 cells were seeded at equal density into 96-well cell culture plates. After 48 hours the plates were analyzed with phase contrast microscopy with 20 x and 40 x microscope objectives (CKX41, Olympus).

Illumination of the imaging area

Microscopic studies were performed to compare the illumination at the edge of individual wells in 96-well cell culture plates. Therefore adipose-derived mesenchymal stem cells (AdMSCs) were seeded at equal density into 96-well plates. After 5 days of cell growth, cells were analyzed first with phase contrast microscopy (CK40, Olympus) and, following DAPI (nuclei) staining, with fluorescence microscopy (Life Technologies EVOS®FL, Thermo Fisher Scientific).

Results and Discussion

Planarity and focus stability

Focus stability across the cell growth area is associated with the planarity of the consumable. An enhanced planarity results in less refocusing. As shown in figure 1 the Eppendorf dish displays a clear “in focus” high-contrast picture of the cell layer at all three observation points in the dish. Nearly no refocusing is needed whereas the tested cell culture dishes from other suppliers display a blurred “out of focus” picture of the cells when moving away from the central focus point without a readjustment of the microscope objective. The good focus stability indicates the enhanced bottom planarity of the Eppendorf Cell Culture Consumables.

The optical contact scanning analysis shows that the Eppendorf 96-Well Cell Culture Plate has the best whole plate planarity of all analyzed cell culture plates (see figure 2).

Meniscus formation and phase contrast performance

Figure 3 shows that in the Eppendorf Cell Culture Plate, 96-Well the liquid meniscus is reduced to a minimum. The liquid meniscus is a result of the tissue culture treatment of cell culture consumables, which makes the usually hydrophobic polystyrene hydrophilic. For the Eppendorf tissue culture (TC) treated products the standard, well established technique, of a so called corona discharge is used. This process generates charged groups on the usually hydrophobic polystyrene surface. This facilitates cell attachment and spreading of most adherent cells. But unlike most products from other suppliers, a specific technique is used which limits the TC treatment to the well bottom exclusively. Therefore the sides of the walls remain hydrophobic. This leads to the reduced liquid meniscus and improved optical results observed in Eppendorf Cell Culture Consumables.

To evaluate the effect of the meniscus on the phase contrast image performance, cells were seeded in 96-well cell culture plates and analyzed with phase contrast microscopy using 20 x and 40 x objectives with an inverted microscope.

As shown in figure 4 an improved phase contrast can be achieved within the Eppendorf plate at all magnifications. With both magnifications single cells can be clearly observed with high contrast in the Eppendorf plate in comparison to similar plates from other suppliers where cells appear poor in contrast and hard to focus. The liquid meniscus results in optical interference disturbing the phase contrast effect (see figure 5a). The absence of a liquid meniscus prevents the light refraction leading to an even background illumination and an unaffected image contrast (see figure 5b). The result is an excellent phase contrast throughout the observation area.

Illumination of the imaging area

Phase contrast and fluorescence microscopy studies were performed to compare illumination at the edge of individual wells within each 96-well cell culture plate. In the Eppendorf plate the wells show no shadow that interferes with imaging or visual observation. In contrast the wells from both analyzed competitor plates display a black circular shadow limiting the analyzing area and thus interfering with both light and fluorescence microscopy observations (see figure 6). The uniform illumination of the well is associated with the reduced liquid meniscus and the precise well geometry of the Eppendorf Cell Culture Plate.

Conclusion

An efficient microscopic analysis is crucial in cell culture research, from basic routine check of cell morphology to reliable cell-based assay analysis. Eppendorf Cell Culture Consumables show an optimized microscopic performance:

The excellent bottom planarity results in minimal refocusing facilitating a quick work procedure. The exact well design and the re- duced meniscus enable a uniform illumination without disturbing shadows at the edges and an improved phase contrast over the whole observation area. Because of these properties Eppendorf Cell Culture Consumables are best suited for microscopic applications.

Literature

  1. E, Zantl, R. Phase-contrast light microscopy of living cells cultured in small volumes. Microscopy and Analysis. May 2006; 5–7.
  2. J.A. American biotechnology laboratory. 1989 7(1):8–16.



Ines Kristina Hartmann
Eppendorf AG Barkhausenweg 1 22339 Hamburg
E-Mail: hartmann.i@eppendorf.de


Wolf Wente
Eppendorf AG Barkhausenweg 1 22339 Hamburg

Linda Knauer
Eppendorf Polymere
Barkhausenweg 1 22339 Hamburg

Khandaker Siddiquee
Eppendorf Inc.
Enfield, CT, USA

Ma Sha
Eppendorf Inc.
Enfield, CT, USA